PRODUCT SPECIFICATIONS
PRETRAN-DP-200
Smart differential transmitter
Smart differential transmitter, micropressure, for micropressure settings
01
Classification: Smart differential transmitter
02
Check configuration based on field work
TECHNICAL SPECIFICATIONS
Technical specifications
Check the specifications against your operating conditions and use the selection tables to confirm your order configuration.
View selection tables ↓| Model | PRETRAN-DP-200 |
|---|---|
| Measurement principle | Monocrystalline silicon piezoresistive Micropressurizers + Digital signal processing (smart) |
| Measurement type | Micro differential pressure (D) / Micropressure (G) |
| Measurement range | 0–25Pa to 0–10kPa (turndown ratio 50:1 Adjustable) |
| Minimum span | 25Pa (lower bound to 0–10Pa) |
| Accuracy class | ±0.1% / ±0.2% / ±0.5% (Optional) |
| Maximum static pressure | 2.5MPa / 5MPa (Optional) |
| Overload capacity | 3 Range |
| Output signal | 4–20mA (DC) + HART / RS485 (MODBUS) Optional |
| Power supply | 12–36VDC (standard 24VDC) |
| Load capacity | (Supply voltage − 12 V) / 0.02 A, up to 1.2 kΩ |
| Sensor capsule | 316L (standard), Hastelloy C276 (Optional) |
| Process connection | 1/8NPT / M20×1.5 / φ6 Stainless steel pressurized interface |
| Display type | 4 bit LCD local Display (percentage/project volume) rotates the watch header |
| Damping time | 0–32s Configurable |
| Stability | ±0.2% /year |
| Process medium temperature | -20~+100℃ |
| Ambient temperature | -40~+85℃ |
| Ingress and explosion protection | IP66/67; Ex d II C T6 Gb / Ex ia II C T4 Ga (Optional) |
| Zero shift | Positive/negative migration 100%, free of magnetic pen |
| Applicable standards | GB/T 17614 / IEC 61010 / HART 7 |
| Suitable media | Clean gas micro- differential pressure: Clean room pressure differential, filter jam monitoring, flue flow, HVA C |
Product selection
PRETRAN-DP-200| Code | Range options (Pa) |
|---|---|
| R1 | 0~25 / 0~50 |
| R2 | 0~100 / 0~250 |
| R3 | 0~500 / 0~1000 |
| R4 | 0~2500 / 0~5000 |
| R5 | 0~10000 |
| Code | Accuracy class |
|---|---|
| E1 | ±0.5% |
| E2 | ±0.2% |
| E3 | ±0.1% |
| Code | Maximum static pressure |
|---|---|
| P1 | 2.5MPa |
| P2 | 5MPa |
| Code | Output signal |
|---|---|
| O1 | 4~20mA + HART |
| O2 | 4~20mA + RS485(MODBUS-RTU) |
| Code | Sensor capsule materials |
|---|---|
| M1 | 316L Stainless steel |
| M2 | Hastelloy C276 |
| Code | Process connection |
|---|---|
| J1 | 1/8NPT pressurized interface |
| J2 | M20×1.5 |
| J3 | φ6 carat (Stainless steel baroque) |
| Code | Display and adjustment |
|---|---|
| L1 | No display |
| L2 | 4 LCD Display |
| L3 | LCD + Magnetic pen correction |
| Code | Explosion protection marking |
|---|---|
| X1 | Non-Ex |
| X2 | Flameproof Ex d II C T6 Gb |
| X3 | Intrinsically safe Ex ia II C T4 Ga |
Confirm specifications and ordering options