Pressure (vacuum) calibrationClick on the picture to zoom in.

PRODUCT SPECIFICATIONS

VISCON-PC-100

Pressure (vacuum) calibration

Pressure (vacuum) calibration, pressure/vacuum calibration

01

Classification: Pressure (vacuum) calibration

02

Check configuration based on field work

⁇ Return pressure (vacuum) calibration classification

TECHNICAL SPECIFICATIONS

Technical specifications

VISCON-PC-100

Check the specifications against your operating conditions and use the selection tables to confirm your order configuration.

View selection tables ↓
ModelVISCON-PC-100
Operating principleManual piston pump Vacuum, pressure of cavity measured by high Accuracy pressure sensitivity component (Diffused silicon / Monocrystalline silicon) and numerically Display, with fine-tune valves and excretion valves accurately stabilizing to the point of detection, direct on-site verification of the inspected pressure sheet, pressure Transmitter, pressure switch
StructureHand pumps (pressure/pump Vacuum double trip), pressure cavities, fine-modified valves, excise pressure valves, precise digital pressure sheets, measured interfaces, electrometric interfaces, all integrated into portable aircraft boxes Integral
Pressure generation methodManual pressurization: piston pump presses continuously to target value, fine-modified valves, slow release of depressor valves; manual; pump vacuum: switch to re-suction to -95kPa (Gauge pressure)
Pressure rangeVacuum Type -95kPa–0;pressure type -95kPa–0–2.5MPa (optional 0–4MPa); hydraulic type 0–60MPa (The cups of oil, optional 0–100MPa)
Calibration rangeVacuum pressure composite table, pressure table, pressure transmitter and pressure switch that covers -0.1–0–2.5MPa; hydraulic type 0–60MPa high pressure table
Accuracy class0.05%FS (standard), 0.1%FS, 0.02%FS (High-accuracy type, only air pressure/ Vacuum Range)
Pressure stability1min internal drift ≤0.05%FS; voltage 5min pressure ≤0.1%FS (hythmic and sealed under satisfactory conditions)
Vacuum levelLimits vacuum ≤5kPa absolute pressure (About -95kPa gauge pressure); vacuum Rate ≤30s From constant pressure to -90kPa
Display type5 " Colour TFT LCD, double row display (Main value pressure + by-product/percentage/peak); backlight, visible in the sun
Display unitskPa, MPa, bar, psi, kgf/cm2, mmHg, mmH2O, inHg Wait. 10 Seeds, one key switch
Electrical signal measurementDirect current 0–25.000mA (accuracy 0.05%RD+0.01%FS); built-in DC 24V Loop power supply (with short-circuit protection), which simultaneously provides power to Transmitter and reads output currents
Calibration functionsZero dot decorating, peak/valve maintenance, pressure check leak (pressure metering and pressure reduction rate), pressure switch up/down switch automatic capture and switch calculation, Transmitter 4 / 5 Linearity error automatic calculation
Working mediumAir pressure/ vacuum: Clean air or nitrogen gas (prohibited oil); hydraulic type: transformer oil or pure water (optional 40# hydraulic fluid)
Wetted materials316L stainless steel Console and cavity standard); seals NBR NBR (standard), optional FKM FKM or perfluoroethers FFKM (Corrosion resistant Media)
Process connectionStandard Table interface M20×1.5 female thread + Checked interfaces M20×1.5 female thread (each) 1 (a) Roads; optional G½, G¼, 1/4NPT or DN16 Quick lock-in, matchable to switch-to-back group
Data managementInternal storage ≥1000 group check record (with time stamp, range, error) supported U dispatch CSV/Excel optional bluetooth upload phone APP
Communication interfaceRS232 / USB (standard); optional RS485, Bluetooth 5.0
Power supplyChargeable lithium batteries 7.4V / 5200mAh, working continuously ≥15h; equipped with AC 220V adapter to support side-loading
Operating environmentTemperature 0–50℃, relative humidity ≤90%RH; storage temperature -20–+70℃; no powerful vibration and corrosive gases
Ingress protectionIP54 (standard); optional IP65
Overall dimensionsAbout 260 × 150 × 78mm (including handbands and connections, without hoses)
Instrument weightAbout 1.6kg (pressure/ vacuum type); about 2.4kg (oiled cups)
Applicable standardsJJG 875 Digital Pressure Sheets/ JJG 52 "Flexible Widget General Pressure Sheet, Pressure" vacuum Table and vacuum Table / JJG 49 "Flexible Widget Precision Pressure Sheets and vacuum Table / JJG 882 "Pressure" transmitter 》/ GB/T 17626 (EMC)
Typical applicationsOn-site verification of power plants, petrochemicals, metallurgicals, water treatment, etc. Vacuum gauge, pressure Vacuum table, pressure Transmitter, pressure switch online; rapid comparison of mass pressure tables in instrument workshops; field and high-altitude operations with no external gas, no electricity
Product featuresA device combines pressure construction and measurement with no additional pressure source; vacuum with positive pressure double-direction construction, a cover -95kPa–2.5MPa complex range; fine-tune valves, single point stabilization time ≤30s; inner 24V Loop power supply with current measurement, Transmitter check one connection completed; full Chinese menu, with gloves to operate

Product selection

VISCON-PC-100
Pressure generation type
CodePressure generation type
V1Vacuum (-95kPa–0)
V2Pressure type (-95kPa–0–2.5MPa, Vacuum positive-pressure dual function)
V3Hydraulic type (0–60MPa)
V4Pressure + hydraulic double configuration
Pressure range
CodePressure range
R1-95–0kPa (Vacuum)
R2-95kPa~0~600kPa
R3-95kPa~0~1.6MPa
R4-95kPa–0–2.5MPa (Standard)
R5-95kPa~0~4MPa
R60–60MPa (prime pressure)
R70–100MPa (prime pressure, order)
Accuracy class
CodeAccuracy class
A10.02%FS (High-accuracy type)
A20.05%FS (Standard)
A30.1%FS
Function Configuration
CodeFunction Configuration
F1Base type (pressure + pressure measurement)
F2Pressure + 4–20mA electron (include 24V Loop power supply)
F3F2 + pressure switch test
F4F3 + Verify Data Records and Export
Process connection
CodeProcess connection
T1M20×1.5 Female thread (Standard, two-way)
T2G½ Female thread
T31/4NPT Female thread
T4DN16 fast lock-in
Seals and wetted materials
CodeSeals and wetted materials
M1NBR Sealed + 316L interface (Standard, air/nitrogen)
M2FKM sealed + 316L (temperature and oil)
M3FFKM Pfluorinated Ether + 316L (Forced corrosive medium)
M4Hydraulic type (oil cup + 316L + FKM)
Communication and storage
CodeCommunication and storage
C1RS232 + USB (Standard)
C2C1 + RS485
C3C1 + Bluetooth 5.0 (cell phone APP)
Power supply and protection
CodePower supply and protection
B1Lithium Battery + AC adaptor, IP54
B2Lithium Battery + AC adaptor, IP65
B3Large capacity batteries (continuation ≥25h), IP65

Confirm specifications and ordering options

Contact us with the measurement target, operating conditions, and installation requirements to confirm your configuration.

Contact us