Pressure CheckerClick on the picture to zoom in.
PRODUCT SPECIFICATIONS
VISCON-PC-150
Pressure Checker
Pressure checker, desktop checker.
01
Classification: Pressure Checker
02
Check configuration based on field work
TECHNICAL SPECIFICATIONS
Technical specifications
Check the specifications against your operating conditions and use the selection tables to confirm your order configuration.
View selection tables ↓| Model | VISCON-PC-150 |
|---|---|
| Operating principle | The pressure is stabilized by the pressure-making system within the closed circuit, the pressure is accurately stabilized at each control point by the steady pressure valve and the fine-modified valve, and is also sent to multiple inspected watch positions simultaneously through the distribution route; the index Standard (or piston pressureometer) installed on the surface is used as a point-by-point comparison to the value Standard for reading the sample and completing the mass pressure sheet check/correction |
| Structure | Pressure pumps, pressure cavities, pressure valves, fine-tune valves, excretion/return valves, oil-water separators and filters, Stainless steel distribution line, multi-position Test gauge mount, table Standard table interface, oil and storage drawers, counters |
| Calibration medium | Hydraulic type: transformer oil (25#) or pure water (Optional ethanol water solubility, anti-frozen); hydraulic type: clean air or nitrogen gas (no oil); two-media type: oil, gas, two separate tubes, toggle valve switching |
| Pressure range | Hydraulic 0–6MPa / 0–25MPa / 0–60MPa (optional 0–100MPa); air pressure -0.095–0–1.6MPa / 0–4MPa / 0–6MPa; bimedia cover -0.095–0–60MPa |
| Pressure generation method | Manual spiral pressure pump (liquid pressure, Standard), manual piston pump (air pressure), electric piston pump (liquid pressure, Optional), aerodynamic booster pump (air pressure, Optional); both finely fined valves |
| Pressure stability | After steady pressure 1min ≤0.05%FS; 10min ≤0.1%FS; system sealability: voltage 10min pressure ≤0.5%FS (under rated pressure) |
| Number of test stations | Standard 3 workspaces (including 1 standard table) or 6 workspaces; optional 8 workspaces, 12 workspaces; individual cut-off valves for each workspace |
| Test gauge mount | M20×1.5 standard thread watch (replaceable G½, G¼, 1/4NPT, M14×1.5 conversion interface); table adjusts upward and downwards to Φ60–Φ250 and surface 30–90mm |
| Clamping method | Manual thread Tighten (standard); optional Aerodynamic Quick Tighten (Teach/button Control, Single Tighten ≤3s) |
| Reference gauge configuration | 0.05 Classified Digital Pressure Sheet (standard); optional 0.02 Digital Pressure Sheet, 0.4/0.25 Class Precision Pointer Table, 0.05 Pistol Pressure Sheet; Standard Table can be removed for separate screening |
| Pipes and valves | 316L stainless steel seamless pipe, high pressure cut-off valve, steady pressure valve, fine-modified valve, depressure valve, refuelling valve; with 5μm filter and oil water separators to prevent impurities from entering the surface |
| Transmitter calibration module | Optional _Other Organiser DC 24V loop power supply + 0–25mA Digital current meters (DMP) 0.05% RD) To achieve pressure transmitter Synchronization of “pressure-reading currents”; optional 4 Roads/ 8 Clear the road. Display |
| Bench construction | Powder coating (standard, ash); optional 304 Stainless steel surface and perimeter; Stainless steel oil tank and removable tarpaulin with a road shutter on the back |
| Display and controls | Standard Table display (backlight); optional with 7 inch touch screen concentration Display each post pressure and current, and access to the upper machine software records |
| Data management | Optional System verification software: enter checked information, generate check points by protocol, auto-record Standard values/screened values, calculate errors in the log and return error, determine and print the check records and certificates |
| Power supply | AC 220V ±10%, 50Hz; whole power ≤1.5kW (electrical pump type ≤2.2kW) |
| Operating environment | Temperature 5–35℃, relative humidity ≤85%RH; placed on the ground level away from vibrating sources and corrosive gases |
| Overall dimensions | About 1500 × 750 × 1250mm (6 place of work standard table); increase in the number of hours worked by the number of hours on the table at 200mm /space |
| Instrument weight | Hydraulic type approximately 180kg; air pressure type about 140kg; two-media type about 220kg |
| Applicable standards | JJG 52 "Flexible Widget General Pressure Sheet, Pressure" vacuum Table and vacuum Table / JJG 49 "Flexible Widget Precision Pressure Sheets and vacuum Table / JJG 875 Digital Pressure Sheets/ JJG 882 "Pressure" transmitter 》/ JJG 544 Pressure controllers/ GB/T 17626 (EMC) |
| Typical applications | Measurement control posts, pressure instrument manufacturing plant finished line, power plant and petrochemical equipment workshop, pressure gauge batch overhaul and periodic check-up; single desk meets the full daily stress sheet control workload of a class |
| Product features | Pressure, pressure, distribution, benchmark, reading is concentrated in one, and operators sit down to complete all calibration actions; multi-position multiple meters, mass readings at the same point, efficiency increases more than 5 for individual verifications; steady pressure valve + micro-modified valves steady double pressure, single point stabilization time ≤30s; Standard table can be removed for review without affecting the use of the table; oil compactors and filters reduce medium pollution and maintain low volumes |
Product selection
VISCON-PC-150| Code | Calibration medium |
|---|---|
| L1 | Hydraulic (voltage transformer oil) |
| L2 | Hydraulic pressure (purified water/ethanol water solution) |
| L3 | Air pressure (air/nitrogen) |
| L4 | Double media (oil, gas independent pipeline) |
| Code | Pressure range |
|---|---|
| R1 | 0–6MPa (prime pressure) |
| R2 | 0–25MPa (prime pressure) |
| R3 | 0–60MPa (prime pressure, Standard) |
| R4 | 0–100MPa (prime pressure, order) |
| R5 | -0.095–0–1.6MPa (air pressure) |
| R6 | -0.095–0–6MPa (air pressure) |
| R7 | Double medium -0.095–0–60MPa |
| Code | Number of test stations |
|---|---|
| N1 | 3 Workspace (including 1 Standard Table) |
| N2 | 6 Workspace (Standard) |
| N3 | 8 |
| N4 | 12 |
| Code | Pressure generation method |
|---|---|
| P1 | Manual spiral pressure pump (liquid pressure) |
| P2 | Manual piston pump (air pressure) |
| P3 | Electric cylinder plug pump (liquid pressure, button/tamp control) |
| P4 | Aerodynamic booster pump (air pressure, air pressurizer interface) |
| Code | Reference gauge configuration |
|---|---|
| S1 | 0.05 Classified digital pressure table (Standard) |
| S2 | 0.02 Classified digital pressure sheet |
| S3 | 0.4 accurate table of pointers |
| S4 | 0.05 Class Pistol pressureometer |
| Code | Clamping method |
|---|---|
| C1 | Manual Thread Top (Standard) |
| C2 | It's moving fast. |
| C3 | Aerodynamic fast-strength (button, place-by-place) |
| Code | Bench configuration |
|---|---|
| T1 | Powder coating steel table (Standard) |
| T2 | 304 Stainless steel stage and perimeter |
| T3 | T2 + transmitter calibration module (24V + current meter) |
| T4 | T3 + 7 " Touch screen focus Display |
| Code | Software and records |
|---|---|
| W1 | Non-enabled software (manual recording) |
| W2 | Basic version (checking log sheet + certificate printing) |
| W3 | Standard version (Protocol library + error determination + account) |
| W4 | Web-based version (database/multi-end, relay measurement management system) |
Confirm specifications and ordering options